PurposeTo describe recent changes in the dimensions of the ERG-jet contact lens electrode, and to suggest corresponding changes in the dimensions of cylindrical stents to facilitate the use of ERG-jet contact lens electrodes in infants and small children.MethodsUpon the observation that existing cylindrical stents no longer fit newly purchased ERG-jet contact lenses, the spacings between the bumps on the front surface of old and new electrodes were measured with calipers, and prototype stents were newly machined from acrylic plastic.ResultsCylindrical stents with an outside diameter of 0.277inches (about 7.04mm) were found to fit snugly between the front surface bumps of new ERG-jet contact lens electrodes.ConclusionsRecent modifications in the injection molds for ERG-jet contact lens electrodes have resulted in an increase in the space between the bumps on the front surface of the electrodes, necessitating a corresponding increase in the outside diameter of cylindrical stents intended to fit snugly between the bumps for ease of insertion and prevention of lid closure.