期刊
BIOMEDICAL OPTICS EXPRESS
卷 4, 期 7, 页码 1051-1060出版社
OPTICAL SOC AMER
DOI: 10.1364/BOE.4.001051
关键词
-
A photonic integrated microfluidic device is demonstrated to perform optical excitation and forward scatter collection all on-chip in a planar format. Integrated on-chip optics formed a tailored beam geometry for optimal excitation of particles while a special design modification allowed for on-chip forward collection with the beam shaping capabilities. A notch was placed in the lens system that caused a dark spot on the facet of a collection waveguide while not affecting the beam geometry at the point of interrogation. The modified device with the ability to form a 10 mu m beam geometry was demonstrated to detect the forward scatter from blank 5 mu m diameter polystyrene beads. Free-space collection of side scatter signals was performed simultaneously with the on-chip collection and the designs demonstrated and enhanced SNR while the reliability of detection was determined to be appropriate for many applications. Excellent performance was confirmed via a false positive rate of 0.4%, a missed events rate of 6.8%, and a coincident rate of 96.3% as determined between simultaneously performed free-space and on-chip detection schemes. (C) 2013 Optical Society of America
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐