4.5 Article

WHAT IS THE PRIMARY SOURCE OF DISCORDANCE IN STRAIN MEASUREMENT BETWEEN VENDORS: IMAGING OR ANALYSIS?

期刊

ULTRASOUND IN MEDICINE AND BIOLOGY
卷 39, 期 4, 页码 714-720

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.ultrasmedbio.2012.11.021

关键词

Strain; Myocardial deformation; Inter-vendor difference; Adult

资金

  1. National Institutes of Health [2R44HL071379-04]

向作者/读者索取更多资源

Strain imaging quantifies myocardial deformation in both overt and subclinical myocardial diseases. Strain values have been shown to vary between vendors. The aim of this study was to investigate whether this variation reflects differences of image characteristics or analysis software. We enrolled 45 patients (age 54 +/- 14 years, 26 males) who underwent two echocardiograms on the same day with two different commercially available ultrasound systems. Three apical views were analyzed by two vendor-specific and two digital imaging and communications in medicine (DICOM)-based software packages. The results were compared in two ways; comparison of global longitudinal strain (GLS) by nine different combinations of ultrasound systems and strain software, and comparison of strains by wall, view and global levels. Comparison of GLS between two vendor-specific software packages yielded poor correlation (rho = 0.35). The use of the same software on the images from different vendors was concordant (rho = 0.74 and 0.64; bias = 0.0 and -0.5). There were no significant differences in correlation coefficients among the comparisons with DICOM-based software. There was no significant difference in correlation coefficients among three views (4CV vs. 3CV, p = 0.44; 4CV vs. 2CV, p = 0.47). Comparison of walls showed the septum to have the best correlation (rho = 0.73), and the posterior wall (rho = 0.31; p = 0.005 vs. septum) had the worst. Use of the same software to measure strain in images from different vendors minimizes the variation of GLS. Post-processing is the most important determinant in inter-vendor variation, with differences in acquisition having a small effect. These findings should be examined and confirmed with other combinations of ultrasound machines and deformation software. (E-mail: tom. marwick@utas.edu.au) (C) 2013 World Federation for Ultrasound in Medicine & Biology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据