Imaging of radiation-sensitive samples in transmission electron microscopes equipped with Zernike phase plates

期刊

ULTRAMICROSCOPY
卷 108, 期 2, 页码 126-140

出版社

ELSEVIER
DOI: 10.1016/j.ultramic.2007.03.008

关键词

zernike phase plate; boersch lens; radiation damage; transmission electron microscope; spherical aberration; high-resolution TEM; biological TEM; single molecule TEM; low-voltage TEM

向作者/读者索取更多资源

We have optimized a bright-field transmission electron microscope for imaging of high-resolution radiation-sensitive materials by calculating the imaging dose no needed to obtain a signal-to-noise ratio (SNR) = 5. Installing a Zernike phase plate (ZP) decreases the dose needed to detect single atoms by as much as a factor of two at 300 kV. For imaging larger objects, such as Gaussian objects with full-width at half-maximum larger than 0.15nm, ZP appears more efficient in reducing the imaging dose than correcting for spherical aberration. The imaging dose no does not decrease with extending of chromatic resolution limit by reducing chromatic aberration, using high accelerating potential (U-0 = 300 kV), because the image contrast increases slower than the reciprocal of detection radius. However, reducing chromatic aberration would allow accelerating potential to be reduced leading to imaging doses below 10e(-)/angstrom(2) for a single iodine atom when a CS-corrector and a ZP are used together. Our simulations indicate that, in addition to microscope hardware, optimization is heavily dependent on the nature of the specimen under investigation. (c) 2007 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据