An automated solid phase extraction coupled with electrothermal atomic absorption spectrometric determination of Pb(II) in high salt content samples

期刊

TALANTA
卷 86, 期 -, 页码 128-132

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.talanta.2011.08.045

关键词

Electrothermal atomic absorption spectrometry; Lead ions; Automated system; High salt content samples; W-Rh coated furnace

资金

  1. Unit of Scientific Research Projects of Erciyes University [EUBAP-FBD-09-713]

向作者/读者索取更多资源

A novel Pb(II) analysis in dialysis concentrates and seawater samples has been developed with on-line separation and preconcentration system coupled with electrothermal atomic absorption spectrometer (Fl-ETAAS). Lewatit Monoplus TP207 imminodiacetat chelating resin was used for separation and preconcentration of Pb(II) ions at pH 4.0. The influence of the analytical parameters such as type of eluent, flow rate of eluent and sample, eluent volume and matrix ions were investigated. W-Rh coated furnace was used as the atomization site. A preconcentration factor of 14 and a detection limit (3s/b) of 12 ng L-1, along with the sampling frequency of 21 h(-1) were achieved with a 170 s sample loading time and with 2.8 mL sample consumption. The relative standard deviation (RSD) was 1.6% for 1 mu g L-1 Pb(II) level. The developed method was used for Pb(II) analysis in dialysis concentrates and seawater samples. The certified reference material (CRM403) experimental results are in good agreement with the certified value. (C) 2011 Elsevier B.V. All rights reserved.