期刊
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY
卷 65, 期 2, 页码 171-175出版社
PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.sab.2010.01.003
关键词
Polyaniline; Mercury; Electrochemical cold vapor generation; Atomic fluorescence spectrometry
类别
资金
- National Nature Science Foundation of PR China [20675074]
An electrochemical cold vapor generation system with polyaniline modified graphite electrode as cathode material was developed for Hg (II) determination by coupling with atomic fluorescence spectrometry. This electrochemical cold vapor generation system with polyaniline/graphite electrode exhibited higher sensitivity: excellent stability and lower memory effect compared with graphite electrode electrochemical cold vapor generation system. The relative standard deviation was 2.7% for eleven consecutive measurements of 2 ng mL(-1) Hg (II) standard solution and the mercury limit of detection for the sample blank solution was 1.3 pg mL(-1) (3 sigma). The accuracy of the method was evaluated through analysis of the reference materials GBW09101 (Human hair) and GBW 08517 (Laminaria Japonica Aresch) and the proposed method was successfully applied to the analysis of human hairs. (C) 2010 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据