期刊
SENSORS AND ACTUATORS B-CHEMICAL
卷 199, 期 -, 页码 154-160出版社
ELSEVIER SCIENCE SA
DOI: 10.1016/j.snb.2014.03.096
关键词
Nitrogen-doped graphene; Poly(3,4-ethylenedioxythiophene); Catechol; Hydroquinone; Electrochemical sensing
资金
- National Natural Science Foundation of China [21103092]
- Fundamental Research Funds for the Central Universities [30920130111003]
- Qing Lan Project
- Science and Technology Support Plan and PAPD of Jiangsu Province, China [BE2011835, BE2013126]
- [NCET-12-0629]
A highly sensitive and selective method was developed for the simultaneous detection of catechol and hydroquinone on a poly(3,4-ethylenedioxythiophene)/nitrogen-doped graphene composite layer. The composite modified electrode was prepared by the electro-polymerization of 3,4-ethylenedioxythiophene on the nitrogen-doped graphene coated glassy carbon electrode. The composite film, characterized by Raman spectra and SEM, was highly efficient to catalyze the redox of catechol and hydroquinone. Two couple of well-defined redox peaks of catechol and hydroquinone could be observed with good reversibility and the enhanced current responses were obtained. Differential pulse voltammetry was used in the simultaneous and quantitative determination of catechol and hydroquinone under the optimized experiment parameters. The proposed sensor based on poly(3,4-ethylenedioxythiophene)/nitrogen-doped graphene exhibited a linear response to the concentrations ranging from 1 to 10 mu M with a low detection limit of 0.26 mu M for catechol and 0.18 mu M for hydroquinone, respectively. (C) 2014 Elsevier B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐