In this study, poly(3,4-ethylenedioxythiophene), PEDOT, was electropolymerized on a sensing chip to entrap glucose oxidase (GOD). The interdigitated array (IDA) electrode and microfluidic channel of the sensing chip was fabricated by photolithography. The IDA electrodes consist of two working electrodes in which one (WE1) was the enzyme-modified electrode and the other was a Pt (WE2) for eliminating noise effect. The microfluidic channel was formed on etched silicon by PDMS (poly(dimethylsiloxane)). In the flow injection analysis, a 0.7V (vs. Ag/AgCl) was set on enzyme electrode to detect the catalytic product, H(2)O(2), and the sensing signal was calibrated using the passed charge rather than the peak current. The linear relationship between the charges and the glucose concentrations, ranging from 1 to 10 mM, was obtained with a sensitivity of 157 mu C cm(-2) mM(-1). Besides, the response time and the recovery time were about 15 and 35-75s, respectively. In real human sample test, the error of single-potential and bi-potential were about 140% and 13%. respectively, comparing to the standard value, indicating that the WE2 can lower the interference effect in this system. (C) 2011 Elsevier B.V. All rights reserved.