Miniaturized Protein Microarray with Internal Calibration as Point-of-Care Device for Diagnosis of Neonatal Sepsis

期刊

SENSORS
卷 12, 期 2, 页码 1494-1508

出版社

MDPI
DOI: 10.3390/s120201494

关键词

protein microarray; streptavidin coated magnetic particles; miniaturization; point-of-care testing; neonatal sepsis; internal calibration

资金

  1. Austrian Research Promotion Agency (FFG) [823977]
  2. Austrian NANO Initiative

向作者/读者索取更多资源

Neonatal sepsis is still a leading cause of death among newborns. Therefore a protein-microarray for point-of-care testing that simultaneously quantifies the sepsis associated serum proteins IL-6, IL-8, IL-10, TNF alpha, S-100, PCT, E-Selectin, CRP and Neopterin has been developed. The chip works with only a 4 mu L patient serum sample and hence minimizes excessive blood withdrawal from newborns. The 4 mu L patient samples are diluted with 36 mu L assay buffer and distributed to four slides for repetitive measurements. Streptavidin coated magnetic particles that act as distinct stirring detection components are added, not only to stir the sample, but also to detect antibody antigen binding events. We demonstrate that the test is complete within 2.5 h using a single step assay. S-100 conjugated to BSA is spotted in increasing concentrations to create an internal calibration. The presented low volume protein-chip fulfills the requirements of point-of-care testing for accurate and repeatable (CV < 14%) quantification of serum proteins for the diagnosis of neonatal sepsis.