4.5 Article

A radial collimator for a time-of-flight neutron spectrometer

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 85, 期 8, 页码 -

出版社

AMER INST PHYSICS
DOI: 10.1063/1.4891302

关键词

-

资金

  1. Scientific User Facilities Division, Office of Basic Energy Sciences, U. S. Department of Energy

向作者/读者索取更多资源

We have engineered and installed a radial collimator for use in the scattered beam of a neutron time-of-flight spectrometer at a spallation neutron source. The radial collimator may be used with both thermal and epithermal neutrons, reducing the detected scattering intensity due to material outside of the sample region substantially. The collimator is located inside of the sample chamber of the instrument, which routinely cycles between atmospheric conditions and cryogenic vacuum. The oscillation and support mechanism of the collimator allow it to be removed from use without breaking vacuum. We describe here the design and characterization of this radial collimator. (C) 2014 AIP Publishing LLC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据