期刊
REVIEW OF SCIENTIFIC INSTRUMENTS
卷 82, 期 5, 页码 -出版社
AMER INST PHYSICS
DOI: 10.1063/1.3589267
关键词
-
资金
- National Science Foundation (NSF)
- National Aeronautics and Space Administration (NASA)
Velocity errors in particle tracking velocimetry (PTV) are studied. When using high-speed video cameras, the velocity error may increase at a high camera frame rate. This increase in velocity error is due to particle-position uncertainty, which is one of the two sources of velocity errors studied here. The other source of error is particle acceleration, which has the opposite trend of diminishing at higher frame rates. Both kinds of errors can propagate into quantities calculated from velocity, such as the kinetic temperature of particles or correlation functions. As demonstrated in a dusty plasma experiment, the kinetic temperature of particles has no unique value when measured using PTV, but depends on the sampling time interval or frame rate. It is also shown that an artifact appears in an autocorrelation function computed from particle positions and velocities, and it becomes more severe when a small sampling-time interval is used. Schemes to reduce these errors are demonstrated. (C) 2011 American Institute of Physics. [doi: 10.1063/1.3589267]
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐