The authors demonstrate readout of electrically detected magnetic resonance at radio frequencies by means of a LCR tank circuit. Applied to a silicon field-effect transistor at millikelvin temperatures, this method shows a 25-fold increased signal-to-noise ratio of the conduction band electron spin resonance and a higher operational bandwidth of > 300 kHz compared to the kilohertz bandwidth of conventional readout techniques. This increase in temporal resolution provides a method for future direct observations of spin dynamics in the electrical device characteristics. (C) 2009 American Institute of Physics. [doi:10.1063/1.3258206]