4.2 Article

Dosimetric characterisation of bismuth shields in CT: measurements and Monte Carlo simulations

期刊

RADIATION PROTECTION DOSIMETRY
卷 133, 期 2, 页码 105-110

出版社

OXFORD UNIV PRESS
DOI: 10.1093/rpd/ncp025

关键词

-

资金

  1. F&L Medical Product and Radcal Corporation

向作者/读者索取更多资源

Although bismuth shields are frequently used in radiology to reduce radiation dose, its mechanism has not been fully investigated. Dosimetric characteristics of bismuth shields in computed tomography (CT) were assessed with ion chamber and Monte Carlo (MC) simulations. Primary attenuation and backscatter effects of paediatric (2-ply) and adult (4-ply) bismuth shields were measured. Simulated CT beams were used for ion chamber measurements. Radiation doses were measured free-in-air and in the tissue-equivalent slabs. MC simulations for the same settings were also performed. Mean dose reductions from primary attenuation were 23% (2-ply) and 40% (4-ply). The dose increase from backscatter was 2% for both shields. MC simulations for primary beam dose reduction were 20% (2-ply) and 38% (4-ply); the backscatter dose increase was around 6% for both shields. In summary, primary attenuation is the major factor that introduces the dose reduction in bismuth and the dose increase from backscatter is negligible.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据