4.6 Article

Effective atomic numbers and electron densities of some human tissues and dosimetric materials for mean energies of various radiation sources relevant to radiotherapy and medical applications

期刊

RADIATION PHYSICS AND CHEMISTRY
卷 102, 期 -, 页码 139-146

出版社

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.radphyschem.2014.04.033

关键词

Effective atomic number; Electron density; Mean photon energy; Radiation source

向作者/读者索取更多资源

Effective atomic numbers, Z(eff), and electron densities, n(eff), are convenient parameters used to characterise the radiation response of a multi-element material in many technical and medical applications. Accurate values of these physical parameters provide essential data in medical physics. In the present study, the effective atomic numbers and electron densities have been calculated for some human tissues and dosimetric materials such as Adipose Tissue (ICRU-44), Bone Cortical (ICRU-44), Brain Grey/White Matter (ICRU-44), Breast Tissue (ICRU-44), Lung Tissue (ICRU-44), Soft Tissue (ICRU-44), LiF TLD-100H, TLD-100, Water, Borosilicate Glass, PAG (Gel Dosimeter), Fricke (Gel Dosimeter) and OSL (Aluminium Oxide) using mean photon energies, E-m, of various radiation sources. The used radiation sources are Pd-103, Tc-99, Ra-226, I-131, Ir-192, Co-60, 30 kVp, 40 kVp, 50 kVp (Intrabeam, Carl Zeiss Meditec) and 6 MV (Mohan-6 MV) sources. The E-m values were then used to calculate Z(eff) and n(eff) of the tissues and dosimetric materials for various radiation sources. Different calculation methods for Z(eff) such as the direct method, the interpolation method and Auto-Z(eff) computer program were used and agreements and disagreements between the used methods have been presented and discussed. It has been observed that at higher E-m values agreement is quite satisfactory (Dif. < 5%) between the adopted methods. (C) 2014 Elsevier Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据