4.5 Article

A motion artifact generation and assessment system for the rapid testing of surface biopotential electrodes

期刊

PHYSIOLOGICAL MEASUREMENT
卷 36, 期 1, 页码 1-25

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0967-3334/36/1/1

关键词

dry electrodes; surface electrodes; ECG; EMG; motion artifact; electrode-skin impedance; testing; prototyping

资金

  1. Finnish Cultural Foundation (Pirkanmaa Fund)

向作者/读者索取更多资源

Dry electrodes can reduce cost while increasing the usability and comfort of wearable monitoring systems. They are, however, susceptible to motion artifacts. The present electrode testing methods lack reliability and do not separate the factors that affect the motion artifact. In this paper, we introduce a first generation motion artifact generation and assessment system that generates the speed, amplitude, and pattern-wise programmable movement of the electrode. The system simultaneously measures electrode-skin impedance, the motion artifact, and one channel of an electrocardiogram that contains the motion artifact and monitors the mounting force applied to the electrode. We demonstrate the system by comparing the applied movement and the measured signals for electrode movements up to 6 mm and movement frequencies from 0.4 Hz to 4 Hz. Results show that the impedance change and surface potential are visually clearly related to the applied motion, with average correlations of 0.89 and 0.64, respectively. The applied force, electrode location, and electrode structure all affect the motion artifact. The setup enables the motion of the electrode to be accurately controlled. The system can be used as a precursor to the testing of integrated systems because it enables thorough, repeatable, and robust motion artifact studies. The system allows a deeper insight into motion artifacts and the interplay of the various factors that affect them.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据