Ultrasound elastic tensor imaging: comparison with MR diffusion tensor imaging in the myocardium

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 57, 期 16, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/57/16/5075

关键词

-

资金

  1. Agence Nationale de la Recherche (ANR)
  2. Societe Francaise de Cardiologie (SFC)
  3. Federation Francaise de Cardiologie (FFC)

向作者/读者索取更多资源

We have previously proven the feasibility of ultrasound-based shear wave imaging (SWI) to non-invasively characterize myocardial fiber orientation in both in vitro porcine and in vivo ovine hearts. The SWI-estimated results were in good correlation with histology. In this study, we proposed a new and robust fiber angle estimation method through a tensor-based approach for SWI, coined together as elastic tensor imaging (ETI), and compared it with magnetic resonance diffusion tensor imaging (DTI), a current gold standard and extensively reported non-invasive imaging technique for mapping fiber architecture. Fresh porcine (n = 5) and ovine (n = 5) myocardial samples (20 x 20 x 30 mm(3)) were studied. ETI was firstly performed to generate shear waves and to acquire the wave events at ultrafast frame rate (8000 fps). A 2.8 MHz phased array probe (pitch = 0.28 mm), connected to a prototype ultrasound scanner, was mounted on a customized MRI-compatible rotation device, which allowed both the rotation of the probe from -90 degrees to 90 degrees at 5 increments and co-registration between two imaging modalities. Transmural shear wave speed at all propagation directions realized was firstly estimated. The fiber angles were determined from the shear wave speed map using the least-squares method and eigen decomposition. The test myocardial sample together with the rotation device was then placed inside a 7T MRI scanner. Diffusion was encoded in six directions. A total of 270 diffusion-weighted images (b = 1000 s mm(-2), FOV = 30 mm, matrix size = 60 x 64, TR = 6 s, TE = 19 ms, 24 averages) and 45 B-0 images were acquired in 14 h 30 min. The fiber structure was analyzed by the fiber-tracking module in software, MedINRIA. The fiber orientation in the overlapped myocardial region which both ETI and DTI accessed was therefore compared, thanks to the co-registered imaging system. Results from all ten samples showed good correlation (r(2) = 0.81, p < 0.0001) and good agreement (3.05 degrees bias) between ETI and DTI fiber angle estimates. The average ETI-estimated fractional anisotropy (FA) values decreased from subendocardium to subepicardium (p < 0.05, unpaired, one-tailed t-test, N = 10) by 33%, whereas the corresponding DTI-estimated FA values presented a change of -10% (p > 0.05, unpaired, one-tailed t-test, N = 10). In conclusion, we have demonstrated that the fiber orientation estimated by ETI, which assesses the shear wave speed (and thus the stiffness), was comparable to that measured by DTI, which evaluates the preferred direction of water diffusion, and have validated this concept within the myocardium. Moreover, ETI was shown capable of mapping the transmural fiber angles with as few as seven shear wave propagation directions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据