Intra-cavitary dosimetry for IMRT head and neck treatment using thermoluminescent dosimeters in a naso-oesophageal tube

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 54, 期 12, 页码 3649-3657

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/54/12/003

关键词

-

向作者/读者索取更多资源

Complex intensity-modulated radiation therapy (IMRT) treatment plans require rigorous quality assurance tests. The aim of this study was to independently verify the delivered dose inside the patient in the region of the treatment site. A flexible naso-gastric tube containing thermoluminescent dosimeters (TLDs) was inserted into the oesophagus via the sinus cavity before the patient's first treatment. Lead markers were also inserted into the tube in order that the TLD positions could be accurately determined from the lateral and anterior posterior electronic portal images taken prior to treatment. The measured dose was corrected for both daily linac output variations and the estimated dose received from the portal images. The predicted dose for each TLD was determined from the treatment planning system and compared to the measured TLD doses. The results comprise 431 TLD measurements on 43 patients. The mean measured-to-predicted dose ratio was 0.988 +/- 0.011 (95% confidence interval) for measured doses above 0.2 Gy. There was a variation in this ratio when the measurements were separated into low dose (0.2-1.0 Gy), medium dose (1.0-1.8 Gy) and high dose (> 1.8 Gy) measurements. The TLD-loaded, naso-oesophageal tube for in vivo dose verification is straightforward to implement, and well tolerated by patients. It provides independent reassurance of the delivered dose for head and neck IMRT.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据