A simple and effective method for validation and measurement of collimator output factors for Leksell Gamma Knife® Perfexion™

期刊

PHYSICS IN MEDICINE AND BIOLOGY
卷 54, 期 12, 页码 3897-3907

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0031-9155/54/12/019

关键词

-

向作者/读者索取更多资源

Gamma Knife radiosurgery. The new Leksell Gamma Knife (R) Perfexion (TM) system has a completely redesigned collimator system and the collimator output factors are different from the other Leksell Gamma Knife (R) models. In this study, a simple method was developed to validate the collimator output factors specifically for Leksell Gamma Knife (R) Perfexion (TM). The method uses double-shot exposures on a single film to eliminate repeated setups and the necessity to acquire dose calibration curves required for the traditional film exposure method. Using the method, the collimator output factors with respect to the 16 mm collimator were measured to be 0.929 +/- 0.009 and 0.817 +/- 0.012 for the 8 mm and the 4 mm collimator, respectively. These values are in agreement (within 2%) with the default values of 0.924 and 0.805 in the Leksell Gamma Plan (R) treatment planning system. These values also agree with recently published results of 0.917 (8 mm collimator) and 0.818 (4 mm collimator) obtained from the traditional methods. Given the efficiency of the method, measurement and validation of the collimator output factors can be readily adopted in commissioning and quality assurance of a Leksell Gamma Knife (R) Perfexion (TM) system.