4.4 Article Proceedings Paper

Customized impedance spectroscopy device as possible sensor platform for biosensor applications

出版社

WILEY-V C H VERLAG GMBH
DOI: 10.1002/pssa.200983305

关键词

biosensor; impedance spectroscopy; DNA; protein; electroanalytical chemistry

向作者/读者索取更多资源

A new impedeance spectroscopy unit is developed fully customized to become a vital part of label free biosensor arrays with possible applications in DNA and protein sensing Test measurements are conducted to expolre the accuracy and specificity of the system, both under electronic lab conditions as well as under wet cell conditions with synthetic-diamond based sensor electrodes. The impedance of seven resistors was monitored for 17h and a maximum error <0.02% was found, Furthermore the impedance of PBS at different concentrations was monitored for 60 min per concentration and a different impedance for each concentration on was detected. The impedance is also monitored for NaOH, PBs and nuclease free water at different temperatures with a total duration of 60 min per fluid. Systematically different impedances for each temperature per fluid were found and the temperature coefficients were determined. All test measurements lead to results well with in specification.(C) 2010 WILEY-VCH Verlag GmbH & Co. KGaA, Weinheim

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据