We present the smallest reported side-viewing needle probe for optical coherence tomography (OCT). Design, fabrication, optical characterization, and initial application of a 30-gauge (outer diameter 0.31mm) needle probe are demonstrated. Extreme miniaturization is achieved by using a simple all-fiber probe design incorporating an angle-polished and reflection-coated fiber-tip beam deflector. When inserted into biological tissue, aqueous interstitial fluids reduce the probe's inherent astigmatism ratio to 1.8, resulting in a working distance of 300 mu m and a depth-of-field of 550 mu m with beam diameters below 30 mu m. The needle probe was interfaced with an 840 nm spectral-domain OCT system and the measured sensitivity was shown to be only 7 dB lower than that of a comparable galvo-scanning sample arm configuration. 3D OCT images of lamb lungs were acquired over a depth range of similar to 600 mu m, showing individual alveoli and bronchioles. (C) 2011 Optical Society of America