4.6 Article

Ultrathin side-viewing needle probe for optical coherence tomography

期刊

OPTICS LETTERS
卷 36, 期 19, 页码 3894-3896

出版社

OPTICAL SOC AMER
DOI: 10.1364/OL.36.003894

关键词

-

类别

向作者/读者索取更多资源

We present the smallest reported side-viewing needle probe for optical coherence tomography (OCT). Design, fabrication, optical characterization, and initial application of a 30-gauge (outer diameter 0.31mm) needle probe are demonstrated. Extreme miniaturization is achieved by using a simple all-fiber probe design incorporating an angle-polished and reflection-coated fiber-tip beam deflector. When inserted into biological tissue, aqueous interstitial fluids reduce the probe's inherent astigmatism ratio to 1.8, resulting in a working distance of 300 mu m and a depth-of-field of 550 mu m with beam diameters below 30 mu m. The needle probe was interfaced with an 840 nm spectral-domain OCT system and the measured sensitivity was shown to be only 7 dB lower than that of a comparable galvo-scanning sample arm configuration. 3D OCT images of lamb lungs were acquired over a depth range of similar to 600 mu m, showing individual alveoli and bronchioles. (C) 2011 Optical Society of America

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据