In-vivo demonstration of a high resolution optical fiber manometry catheter for diagnosis of gastrointestinal motility disorders

期刊

OPTICS EXPRESS
卷 17, 期 6, 页码 4500-4508

出版社

OPTICAL SOC AMER
DOI: 10.1364/OE.17.004500

关键词

-

类别

向作者/读者索取更多资源

Fiber optic catheters for the diagnosis of gastrointestinal motility disorders are demonstrated in-vitro and in-vivo. Single element catheters have been verified against existing solid state catheters and a multi-element catheter has been demonstrated for localized and full esophageal monitoring. The multi-element catheter consists of a series of closely spaced pressure sensors that pick up the peristaltic wave traveling along the gastrointestinal (GI) tract. The sensors are spaced on a 10 mm pitch allowing a full interpolated image of intraluminal pressure to be generated. Details are given of in-vivo trials of a 32-element catheter in the human oesophagus and the suitability of similar catheters for clinical evaluation in other regions of the human digestive tract is discussed. The fiber optic catheter is significantly smaller and more flexible than similar commercially available devices making intubation easier and improving patient tolerance during diagnostic procedures. (C) 2009 Optical Society of America