Diamond dosimetry; Polycrystalline CVD diamond dosimeters at; low bias; Dose and dose-rate under photon and; proton clinical beams
A polycrystalline Chemical Vapour Deposited (pCVD) diamond film has been processed to implement a bidimensional dosimeter for applications in clinical radiotherapy. The detector, with an area of 2.5 x 2.5 cm(2) and a thickness of 300 im, has been equipped with a matrix of electrical contacts (12 x 12 pixels) on the upper surface. The matrix of pixels has been electrically characterized in dark showing uniformity and very low leakage current. This study demonstrates that when applied bias is in the range 0-5 V, the pCVD diamond film provides very accurate measurement of dose and close-rates under clinical 62 MeV proton beam, conventional 6-10 MeV photon beams, and Intensity Modulated Radiotherapy (WIRT) beams. (C) 2013 Elsevier B.V. All rights reserved.