Error in dynamic spring constant calibration of atomic force microscope probes due to nonuniform cantilevers

期刊

NANOTECHNOLOGY
卷 22, 期 29, 页码 -

出版社

IOP Publishing Ltd
DOI: 10.1088/0957-4484/22/29/295703

关键词

-

向作者/读者索取更多资源

Many common atomic force microscope (AFM) spring constant calibration methods regard the AFM probe as a uniform cantilever, neglecting the tip mass and any nonuniformity in the thickness of the probe along its length. This work quantifies the error in the spring constant estimated by the Sader and thermal calibration methods due to nonuniformity in the thickness of the cantilever and the influence of the mass loading effect of the probe tip. Formulae are presented that can be used to compute the uncertainty in cantilever calibration for an arbitrary thickness nonuniformity, or to correct the calibration methods if the thickness nonuniformity is known. The results show that both methods are quite sensitive to nonuniformity. When the first dynamic mode is used in the calibration, the error in the spring constant estimated by either method is between -4% and 9% for a cantilever whose thickness increases or decreases linearly by 30% along its length. The errors are several times larger if the second or higher dynamic modes are used. To illustrate the proposed methods, a commercial AFM probe that has significant nonuniformity is considered and the error in calibrating this probe is quantified and discussed. For this particular probe, variations in the thickness of the probe over the last 15% of its length are found to significantly reduce the accuracy of the calibration when the thermal method is used, since that method is sensitive to changes in the shape of the eigenmode of the probe near its free end.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据