Accuracy of needle position measurements using fiber Bragg gratings

期刊

出版社

TAYLOR & FRANCIS LTD
DOI: 10.3109/13645706.2012.666251

关键词

Needle interventions; FBG; shape estimation; tip position

类别

向作者/读者索取更多资源

Accurate placement of the needle tip is essential in percutaneous therapies such as radiofrequency ablation (RFA) of liver tumors. Use of a robotic system for navigating the needle could improve the targeting accuracy. Real-time information on the needle tip position is needed, since a needle deflects during insertion in tissue. Needle shape can be reconstructed based on strain measurements within the needle. In the current experiment we determined the accuracy with which the needle tip position can be derived from strain measurements using Fiber Bragg Gratings (FBGs). Three glass fibers equipped with two FBGs each were incorporated in a needle. The needle was clamped at one end and deformed by applying static radial displacements at one or two locations. The FBG output was used for offline estimation of the needle shape and tip position. During deflections of the needle tip up to 12.5 mm, the tip position was estimated with a mean accuracy of 0.89 mm (std 0.42 mm). Adding a second deflection resulted in an error of 1.32 mm (std 0.48 mm). This accuracy is appropriate for applications such as RFA of liver tumors. The results further show that the accuracy can be improved by optimizing the placement of FBGs.