4.6 Article

On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA

期刊

MEDICAL PHYSICS
卷 39, 期 6, 页码 3051-3059

出版社

WILEY
DOI: 10.1118/1.4711755

关键词

IMRT; VMAT; 3D-Gamma; EPID; QA

向作者/读者索取更多资源

Purpose: To investigate using 3D c analysis for IMRT and VMAT QA. Methods: We explored and studied 3D c-analysis by comparing TPS computed and EPID backprojection reconstructed doses in patient's CT images. Two 3D c quantities, cPTV and c10, were proposed and studied for evaluating the QA results, and compared to 2D c (MapCheck composite: cMC). Results: It was found that when 3%(global)= 3 mm criteria was used, all IMRT and 90% of VMAT plans passed QA with a c pass rate -90%. A significant statistical correlation was observed between 3D and 2D c-analysis results for IMRT QA if c10 and cMC are concerned, but no significant relation is found between cPTV and cMC. Conclusions: 3D c analysis based on EPID dose back-projection may provide a feasible tool for IMRT and VMAT pretreatment plan QA. (C) 2012 American Association of Physicists in Medicine. [http:/dx.doi.org10.11181.4711755]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据