4.6 Article

Technical Note: Homogeneity of Gafchromic® EBT2 film

期刊

MEDICAL PHYSICS
卷 37, 期 4, 页码 1753-1756

出版社

WILEY
DOI: 10.1118/1.3368601

关键词

dosimeters; dosimetry; radiation therapy

向作者/读者索取更多资源

Methods: A scan protocol established for EBT film and published previously was used. The uniformity of the film coloration was investigated for unirradiated and irradiated EBT2 film sheets. The dose response of EBT2 film was measured and the influence of film inhomogeneities on dose determination was evaluated. Results: Inhomogeneities in pixel values of up to +/- 3.7% within one film were detected. The relative inhomogeneities were found to be approximately independent of the dose. Nonuniformities of the film response lead to uncertainties in dose determination of +/- 8.7% at 1 Gy. When using net optical densities for dose calibration, uncertainties in dose determination amount to more than +/- 6%. Conclusions: EBT2 films from the lot investigated in this study show response inhomogeneities, which lead to uncertainties in dose determination exceeding the commonly accepted tolerance levels. It is important to test further EBT2 lots regarding homogeneity before using the film in clinical routine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据