4.6 Article

On the speckle-free nature of photoacoustic tomography

期刊

MEDICAL PHYSICS
卷 36, 期 9, 页码 4084-4088

出版社

WILEY
DOI: 10.1118/1.3187231

关键词

photoacoustic imaging; speckle

资金

  1. National Institutes of Health [R01 NS46214, R01 EB000712, R01 EB008085, U54 CA136398]

向作者/读者索取更多资源

Purpose: A long-standing conundrum is why photoacoustic tomography (PAT) possesses the unique ability to produce images devoid of speckle artifacts while all other coherent imaging technologies do not. Methods: In this paper, we explain the inherent mechanism that suppresses speckle in PAT, and the analysis was validated by simulations based on an experimental PAT system. Results: We found that the speckle-free feature of PAT results directly from the optical absorption contrast. Conclusions: All optical absorbers expand on laser excitation, and therefore all initial photoacoustic pressure rises are positive, which engenders strong correlations among the photoacoustic waves from the absorbers. As a result, prominent boundaries always build up in photoacoustic images and suppress the interior speckle. (C) 2009 American Association of Physicists in Medicine. [DOI: 10.1118/1.3187231]

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据