Performance of two MALDI-TOF MS systems for the identification of yeasts isolated from bloodstream infections and cerebrospinal fluids using a time-saving direct transfer protocol

期刊

MEDICAL MICROBIOLOGY AND IMMUNOLOGY
卷 203, 期 2, 页码 93-99

出版社

SPRINGER
DOI: 10.1007/s00430-013-0319-9

关键词

MALDI-TOF MS; VitekMS; Biotyper MS; Candida spp.; Yeast; Speed; Identification

资金

  1. Koln Fortune program, Faculty of Medicine, University of Cologne, Germany
  2. program Frankfurter Forschungsforderung (FFF)'', Faculty of Medicine, University of Frankfurt, Germany

向作者/读者索取更多资源

The rapid and correct identification of pathogens is of paramount importance for the treatment of patients with invasive infections. Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) can speed up the identification of bacteria and fungi and has quickly been embraced by medical microbiology laboratories worldwide. Different MALDI-TOF systems have been compared in studies focussing on identification rates of different pathogens. Another aspect that has not been systematically assessed is the performance in daily routine and handling, which is important especially for microbiology routine laboratories. We compared two widespread commercial systems, Microflex LT Biotyper (Bruker) and VitekMS (bioM,rieux), for the identification of 210 relevant clinical yeasts under routine conditions, using a time-saving direct transfer protocol. We assessed the need for an additional extraction step, the threshold for species identification and the duration of measurements with the two systems. The tested yeasts included 34 Candida albicans isolates, 144 non-albicans Candida spp. and 32 yeasts of different genera. The results of the two MS systems were compared with that of biochemical identification and, in case of discrepancies, DNA sequencing of the internal transcribed spacer or the large subunit of ribosomal DNA. Both systems correctly identified 96.2 % of isolates [202/210, non-significant (n.s.)]. Misidentifications were observed for VitekMS only (n = 5, no major errors, n.s.). VitekMS was the slower system (19.8 vs. 8.0 min for 10 samples, p = 0.002) but had the advantage of a more effective direct transfer protocol with less need for an additional extraction step.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据