4.4 Article

New prototype assembly methods for biosensor integrated circuits

期刊

MEDICAL ENGINEERING & PHYSICS
卷 33, 期 8, 页码 973-979

出版社

ELSEVIER SCI LTD
DOI: 10.1016/j.medengphy.2011.03.006

关键词

Biosensor; Integrated circuit; Prototype; Assembly; Packaging; Partial encapsulation

资金

  1. UK Engineering and Physical Sciences Research Council (EPSRC)
  2. University of Bath
  3. Engineering and Physical Sciences Research Council [GR/S08237/01] Funding Source: researchfish

向作者/读者索取更多资源

Two new prototype assembly methods have been evaluated for biosensors that combine an integrated circuit (IC) sensor with a culture chamber. The first method uses a poly-ethylene glycol (PEG) mould to mask the IC sensor during application of a room temperature vulcanising (RTV) silicone elastomer used to insulate the bondpads and bondwires. The second method utilises the 'partial encapsulation' service offered by Quik-Pak, USA. Both methods were shown to provide good electrical insulation and demonstrated biocompatibility with the NG108-15 cell line. These methods are particularly useful for the assembly of low-cost ICs with a small (<4 mm(2)) sensor area. (C) 2011 IPEM. Published by Elsevier Ltd. All rights reserved.