4.5 Article

Portable system for the detection of micromolar concentrations of glucose

期刊

MEASUREMENT SCIENCE AND TECHNOLOGY
卷 25, 期 2, 页码 -

出版社

IOP PUBLISHING LTD
DOI: 10.1088/0957-0233/25/2/025701

关键词

glucose sensor; portable system; fluorescence

资金

  1. National Institutes of Health, National Institutes of Child Health and Human Development [R41HD069207]
  2. GE Healthcare

向作者/读者索取更多资源

Glucose in non-invasively collected biofluids is generally in the micromolar range and thus, requires sensing methodologies capable of measuring glucose at these levels. Here, we present a small fluorometer system that can quantify glucose in the range of 0-5 mu M with resolution of similar to 0.07 mu M. It relies on the glucose binding protein (GBP) fluorescently labeled with two fluorophores. Fluorescence signals from the dual-labeled GBP are utilized in a ratiometric mode, making the measurements insensitive to variations in protein concentration and other systematic errors. Fluorescence is quantified by a miniature, dedicated ratiometric fluorometer that is powered via USB. Concentration is calculated using an ultra-mobile personal computer. The whole system is designed to be pocket sized suitable for point-of-care or bedside applications. Test results suggest that the system is a promising tool for accurate measurements of low glucose concentrations (0.1-10 mu M) in biological samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据