3D printing is a new technology whose applications are only beginning to be explored. In this report, we describe the application of 3D printing to the design and construction of supersonic nozzles. Nozzles can be created for $0.50 or less, and the ease and low cost can facilitate the optimization of nozzle performance for the needs of any particular experiment. The efficacy of a variety of designs is assessed by examining rotational spectra of OCS (carbonyl sulfide) and Ar-OCS using a Fourier transform microwave spectrometer with tandem cavity and chirped-pulse capabilities. A slit geometry which, to the best of our knowledge has not been used in conjunction with Fourier transform microwave spectrometers, was found to increase the signal-to-noise ratio for the J = 1 <- 0 transition of OCS, by a factor of three to four compared with that obtained using our standard circular nozzle. Corresponding gains for the Ar-OCS complex were marginal, at best, but further optimization of nozzle geometry should be possible. The spectrometer itself is designed to allow rapid switching between cavity and chirped-pulse modes of operation without the need to break vacuum. This feature, as well as the newly incorporated chirped-pulse capability, is described in detail. (C) 2015 AIP Publishing LLC.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据