4.3 Article

Application of enzyme/zeolite sensor for urea analysis in serum

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.msec.2014.05.028

关键词

Silicalite; Enzyme adsorption; Urease; Conductometric transducer; Biosensor

资金

  1. Project European IRSES-NANODEVICE (PIRSES-GA) [318524]
  2. National Academy of Sciences of Ukraine

向作者/读者索取更多资源

Urea biosensor based on zeolite-adsorbed urease was applied for analysis of blood serum samples. It should be noted, that this biosensor has a number of advantages, such as simple and fast performance, the absence of toxic compounds during biosensor preparation, high reproducibility and repeatability (RSD = 9% and 4%, respectively). The linear range of urea determination by using the biosensor was 0.003-0.75 mM, and the limit of urea detection was 3 mu M. The method of standard addition was used for analysis of serum samples with 500-fold dilution. Total time of analysis was 10 min. Good reproducibility of urea determination in real samples was demonstrated (RSD = 10%). Biosensor results were verified by using a common method of urea determination (diacetyl monoxime reaction). It was shown that by using this biosensor distinguishing healthy people from people with renal dysfunction becomes easier. (C) 2014 Elsevier B.V. All rights reserved.