4.5 Article

Multichannel receiver coils for improved coverage in cardiac metabolic imaging using prepolarized 13C substrates

期刊

MAGNETIC RESONANCE IN MEDICINE
卷 70, 期 1, 页码 295-300

出版社

WILEY
DOI: 10.1002/mrm.24460

关键词

hyperpolarized 13C; SNR; surface coils; multichannel coils

资金

  1. Natural Sciences and Engineering Research Council of Canada (NSERC PGS-D)
  2. Canadian Institutes of Health Research [CIHR MOP84503]
  3. Heart and Stroke Foundation of Canada [HSF NA7074]

向作者/读者索取更多资源

MR imaging using hyperpolarized 13C substrates has become a promising tool to study real-time cardiac-metabolism in vivo. For such fast imaging of nonrecoverable prepolarized magnetization it is important to optimize the RF-coils to obtain the best signal-to-noise ratio possible, given the required coverage. In this work, three different receiver-coil configurations were computed in pig and human models. The sensitivity maps were demonstrated in phantoms and in vivo experiments performed in pigs. Signal-to-noise ratio in the posterior heart was increased up to 80% with the best multichannel coil as expected. These new coil configurations will allow imaging of the different metabolite signals even in the posterior regions of the myocardium, which is not possible with a single-channel surface-coil. Magn Reson Med, 2013. (c) 2012 Wiley Periodicals, Inc.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据