breast cancer; magnetic resonance spectroscopy; echo-filter; short TE; high spatial resolution
资金
- NCI NIH HHS [P30 CA022453, R21 CA118569, R21 CA 118569-01A1] Funding Source: Medline
The high sensitivity but low specificity of breast MRI has prompted exploration of breast H-1 MRS for breast cancer detection. However, several obstacles still prevent the routine application of in vivo breast H-1 MRS, including poor spatial resolution, long acquisition time associated with conventional multi-voxel MRS imaging (MRSI) techniques, and the difficulty of extra lipid suppression in a magnetic field with relatively poor achievable homogeneity compared to the brain. Using a combination of a recently developed echo-filter (EF) suppression technique and an elliptical sampling scheme, we demonstrate the feasibility of overcoming these difficulties. It is robust (the suppression technique is insensitive to magnetic field inhomogeneity), fast (acquisition time of about 12 min) and offers high spatial resolution (up to 0.6 cm(3) per voxel at 1.5 T with a TE of only 60 ms). This approach should be even better at 3 T with higher resolution and/or shorter TE. (c) 2008 Elsevier Inc. All rights reserved.