4.3 Article

Fracture of titanium nitride-coated femoral component after total knee arthroplasty

期刊

KNEE
卷 21, 期 4, 页码 871-874

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.knee.2014.04.002

关键词

Total knee arthroplasty; Fracture; Femoral component; Titanium nitride coating

向作者/读者索取更多资源

We report a case of fracture of a titanium nitride-coated femoral component 3 years after primary total knee arthroplasty (TKA). The fracture was at the medial condylar area just posterior to the medial peg of the femoral component. The backside of the broken medial condylar portion of the femoral component was devoid of cement. Debonding of the component is a possible cause of the stress fracture. To our knowledge, this is the first case report of the fracture of the femoral component manufactured from titanium alloy. (C) 2014 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据