4.5 Article

Mapping the Stability Diagram of a Quadrupole Mass Spectrometer with a Static Transverse Magnetic Field Applied

期刊

出版社

SPRINGER
DOI: 10.1007/s13361-013-0654-5

关键词

Quadrupole mass spectrometer (QMS); Quadrupole mass filter (QMF); Magnetic field; Zone 1; Lorentz force; Mathieu stability diagram; Ion trajectories; Resolution

资金

  1. Department of Electrical Engineering and Electronics (University of Liverpool)
  2. Engineering and Physical Sciences Research Council (EPSRC)

向作者/读者索取更多资源

Previous experimental and theoretical work identified that the application of a static magnetic (B) field can improve the resolution of a quadrupole mass spectrometer (QMS) and this simple method of performance enhancement offers advantages for field deployment. Presented here are further data showing the effect of the transverse magnetic field upon the QMS performance. For the first time, the asymmetry in QMS operation with B (x) and B (y) is considered and explained in terms of operation in the fourth quadrant of the stability diagram. The results may be explained by considering the additional Lorentz force (v x B) experienced by the ion trajectories in each case. Using our numerical approach, we model not only the individual ion trajectories for a transverse B field applied in x and y but also the mass spectra and the effect of the magnetic field upon the stability diagram. Our theoretical findings, confirmed by experiment, show an improvement in resolution and ion transmission by application of magnetic field for certain operating conditions.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据