3.8 Article

Reliability of the Good Balance System® for Postural Sway Measurement in Poststroke Patients

期刊

JOURNAL OF PHYSICAL THERAPY SCIENCE
卷 26, 期 1, 页码 121-124

出版社

SOC PHYSICAL THERAPY SCIENCE
DOI: 10.1589/jpts.26.121

关键词

Postural sway; Reliability; Stroke

向作者/读者索取更多资源

[Purpose] The purpose of this study was to examine test-retest reliability of the Good Balance system (R) for measurement of postural sway in poststroke patients. [Subjects] Sixty chronic stroke patients (40 men and 20 women; age 63.08 years; stroke duration 16.45 months) participated in this study. [Methods] Postural sway was evaluated using a force platform system (Good Balance system, Metitur Oy, Jyvaskyla, Finland). Two examiners measured postural sway for all participants during two separate testing sessions. The second measurement was performed one week after the first measurement. Intraclass correlation coefficients [ICC(2,1)] were used for estimation of reliability. [Results] The ICC (95% CI) for intra-examiner reliability was good to very good, ranging from 0.69 to 0.93 (0.53-0.96), and the ICC for inter-examiner reliability was good to very good, ranging from 0.85 to 0.98 (0.77-0.99). [Conclusion] The results of the current study indicated that the intra-and inter-examiner reliability of the Good Balance system (R) for measurement of postural sway was good to very good. Therefore, we suggest that measurement of postural sway using the Good Balance system (R) would be useful for clinical assessment in poststroke patients.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据