4.6 Article

Monitoring of adalimumab and antibodies-to-adalimumab levels in patient serum by the homogeneous mobility shift assay

期刊

出版社

ELSEVIER
DOI: 10.1016/j.jpba.2013.01.031

关键词

Adalimumab; Anti-drug antibody; Autoimmune diseases; Therapeutic drug monitoring; Immunogenicity; Mobility shift assay

资金

  1. Prometheus Laboratories, Inc.

向作者/读者索取更多资源

This report describes the analytical validation and application of the homogeneous mobility shift assay (HMSA) method for the measurement of adalimumab and human antibodies-to-adalimumab (ATA) in serum samples from patients who have lost response to adalimumab treatment. Validation of the ATA- and the adalimumab-HMSA revealed a lower limit of detection to be 0.026 U/mL for ATA and 0.018 mu g/mL for adalimumab in serum samples. Intra-assay and inter-assay precision determination yielded a coefficient of variation of less than 15%, and the accuracy of both assays was within 20%. Adalimumab drug tolerance in the ATA-HMSA was up to 20 mu g/mL in the test serum. Serum samples from 100 drug-naive healthy subjects were tested to set-up the cut point of 0.55 U/mL for ATA and 0.68 mu g/mL for adalimumab. Analysis of 100 serum samples from patients who were losing response to adalimumab showed that 26% had an adalimumab level below the cut point, of these 68% were ATA positive. Overall, 44% of the patients (44/100) were positive for ATA. This study presents evidence that drug and anti-drug antibody levels are important determinants of patient response to therapy. (C) 2013 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据