4.7 Article

Design, Validation, and Implementation of a Rapid-Injection NMR System

期刊

JOURNAL OF ORGANIC CHEMISTRY
卷 75, 期 16, 页码 5558-5572

出版社

AMER CHEMICAL SOC
DOI: 10.1021/jo100837a

关键词

-

资金

  1. National Science Foundation (NSF) [CHE-9803124, 0105205, 0414440, CHE 0717989]
  2. Division Of Chemistry
  3. Direct For Mathematical & Physical Scien [0414440, 0105205] Funding Source: National Science Foundation

向作者/读者索取更多资源

A Rapid Injection NMR (RINMR) apparatus has been designed and constructed to allow the observation of fast chemical reactions in real time by NMR spectroscopy. The instrument was designed to allow the rapid (< 2 s) injection and mixing of a metered volume of a reagent into a spinning NMR tube followed by rapid acquisition of the data resulting from the evolution of the chemical process. The various design criteria for this universal system included the ability to deliver any chemical reagent at any temperature and allow for the observation of any nucleus. The various challenges associated with the construction and implementation of this instrument are documented along with the validation of the accuracy of the apparatus with respect to volume and temperature. Finally, the ultimate validation and reproducibility of the technique is presented in the form of three case studies that used the instrument to elucidate various aspects of organic reaction mechanisms. The authors urge interested parties to not embark on the construction of their own instrument and invite those whose research problems might be amenable to this kind of analysis to contact the corresponding author for access to the apparatus described herein.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据