期刊
JOURNAL OF NEUROINTERVENTIONAL SURGERY
卷 7, 期 10, 页码 -出版社
BMJ PUBLISHING GROUP
DOI: 10.1136/neurintsurg-2014-011384.rep
关键词
Aneurysm; Flow Diverter; Stent
Flow diversion is a promising technique for cerebral aneurysm treatment but presents new and sometimes unique challenges. Transient reversible narrowing due to device reconfiguration of the pipeline embolization device (PED) has not been previously described. Here we describe a patient with dolichoectasia of the distal right internal carotid artery with an associated saccular sidewall aneurysm who developed asymptomatic circumferential narrowing of the proximal aspect of the PED in the first month following deployment. The patient was followed conservatively and demonstrated partial resolution of device narrowing on 6 month follow-up imaging.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据