4.3 Article

Dark pixel intensity determination and its applications in normalizing different exposure time and autofluorescence removal

期刊

JOURNAL OF MICROSCOPY
卷 246, 期 1, 页码 1-10

出版社

WILEY
DOI: 10.1111/j.1365-2818.2011.03581.x

关键词

Autofluorescence removal; dark current; dark pixel intensity; exposure time; readout noise; standardization

资金

  1. General Electric Company Global Research Center

向作者/读者索取更多资源

The purpose of this study is to investigate how to scale pixel intensity acquired from one exposure time to another. This is required when comparing grayscale images acquired at different exposure times and other image processing such as autofluorescence removal. Pixel intensity is linear to exposure time as long as images are acquired at the linear range of a camera, but importantly there exists an intercept, which is set by the camera. We termed this intercept as dark pixel intensity, as it is the pixel intensity under conditions of no light and zero exposure time. Dark pixel intensity is determined by camera's readout noise (electron/pixel), gain, and DC offset. Knowing dark pixel intensity, image acquired from one exposure time can be linearly scaled to an image at a different exposure time. Dark pixel intensity can be directly measured by obtaining an image at no light and zero (or minimum) exposure time. It can be also indirectly calculated by capturing images at a series of exposure times. Finally, the prestained and poststained images were acquired at their optimal exposures and autofluorescence was completely removed by normalizing images with the exposure time ratio and dark pixel intensity followed by subtraction.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.3
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据