期刊
JOURNAL OF MICROMECHANICS AND MICROENGINEERING
卷 24, 期 12, 页码 -出版社
IOP PUBLISHING LTD
DOI: 10.1088/0960-1317/24/12/125010
关键词
lead transfer; microassembly; microelectromechanical systems; neural probe array; packaging
类别
资金
- Science and Engineering Research Council of Agency for Science, Technology and Research [132 148 0002]
Brain-machine interface technology can improve the lives of spinal cord injury victims and amputees. A neural interface system, consisting of a 3D probe array and a custom low-power (1 mW) 100-channel (100-ch) neural recording application-specific integrated circuit (ASIC), was designed and implemented to monitor neural activity. In this study, a microassembly 3D probe array method using a novel lead transfer technique was proposed to overcome the bonding plane mismatch encountered during orthogonal assembly. The proposed lead transfer technique can be completed using standard micromachining and packaging processes. The ASIC can be stacking-integrated with the probe array, minimizing the form factor of the assembled module. To minimize trauma to brain cells, the profile of the integrated probe array was controlled within 730 mu m. The average impedance of the assembled probe was approximately 0.55 M Omega at 1 kHz. To verify the functionality of the integrated neural probe array, bench-top signal acquisitions were performed and discussed.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐