4.2 Article

Fetal and maternal heart rate confusion during intra-partum monitoring: comparison of trans-abdominal fetal electrocardiogram and Doppler telemetry

期刊

JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE
卷 25, 期 8, 页码 1517-1520

出版社

INFORMA HEALTHCARE
DOI: 10.3109/14767058.2011.636090

关键词

Doppler telemetry; intra-partum fetal monitoring; labour; trans-abdominal fetal ECG

向作者/读者索取更多资源

Objective: To compare intra-partum performance of transabdominal electrocardiogram with Doppler telemetry. Methods: In this prospective longitudinal study, simultaneous monitoring with trans-abdominal ECG and Doppler telemetry was performed in 41 uncomplicated term singleton pregnancies during labour. Results: The overall success rate for FHR monitoring was similar between trans-abdominal ECG and Doppler telemetry (88.5 +/- 16.7% vs 89.4 +/- 7.6%), except for the second stage of labour. A significantly higher rate of confusion (p < 0.001) between fetal and maternal heart was found for Doppler telemetry (4.5 +/- 4.5%) compared with trans-abdominal ECG (1.3 +/- 1.9%), especially in the second stage and during maternal movements. Conclusions: Trans-abdominal ECG monitoring is feasible, with comparable success rate to traditional Doppler telemetry, without interfering with maternal mobility or requiring midwife intervention. The reduction in maternal\fetal heart rate confusion from trans-abdominal ECG could reduce incorrect obstetric interpretation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据