Validation of Powder X-Ray Diffraction Following EN ISO/IEC 17025

期刊

JOURNAL OF FORENSIC SCIENCES
卷 57, 期 3, 页码 722-737

出版社

WILEY-BLACKWELL
DOI: 10.1111/j.1556-4029.2011.02025.x

关键词

forensic science; powder X-ray diffraction; validation; EN ISO; IEC 17025; laboratory proficiency test; trueness; specificity; selectivity; robustness; preparation errors; detection limit

向作者/读者索取更多资源

Powder X-ray diffraction (PXRD) is used widely in forensic science laboratories with the main focus of qualitative phase identification. Little is found in literature referring to the topic of validation of PXRD in the field of forensic sciences. According to EN ISO/IEC 17025, the method has to be tested for several parameters. Trueness, specificity, and selectivity of PXRD were tested using certified reference materials or a combination thereof. All three tested parameters showed the secure performance of the method. Sample preparation errors were simulated to evaluate the robustness of the method. These errors were either easily detected by the operator or nonsignificant for phase identification. In case of the detection limit, a statistical evaluation of the signal-to-noise ratio showed that a peak criterion of three sigma is inadequate and recommendations for a more realistic peak criterion are given. Finally, the results of an international proficiency test showed the secure performance of PXRD.