4.5 Article

Analyzing Endosonic Root Canal File Oscillations: An In Vitro Evaluation

期刊

JOURNAL OF ENDODONTICS
卷 36, 期 5, 页码 880-883

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2009.12.027

关键词

Endosonics; laser vibrometry; oscillation patterns; ultrasonics

资金

  1. UK Engineering and Physical Sciences Research Council [EP/F020090/1]
  2. EPSRC [EP/F020090/1] Funding Source: UKRI
  3. Engineering and Physical Sciences Research Council [EP/F020090/1] Funding Source: researchfish

向作者/读者索取更多资源

Introduction: Passive ultrasonic irrigation may be used to improve bacterial reduction within the root canal. The technique relies on a small file being driven to oscillate freely within the canal and activating an irrigant solution through biophysical forces such as microstreaming. There is limited information regarding a file's oscillation patterns when operated while surrounded by fluid as is the case within a canal root. Methods: Files of different sizes (#10 and #30, 27 mm and 31 mm) were connected to an ultrasound generator via a 120 degrees file holder. Files were immersed in a water bath, and a laser vibrometer set up with measurement lines superimposed over the files. The laser vibrometer was scanned over the oscillating files. Measurements were repeated 10 times for each file/power setting used. Results: File mode shapes are comprised of a series of nodes/antinodes, with thinner, longer files producing more antinodes. The maximum vibration occurred at the free end of the file. Increasing generator power had no significant effect on this maximum amplitude (p > 0.20). Maximum displacement amplitudes were 17 to 22 mu m (#10 file, 27 mm), 15 to 21 mu m (#10 file, 31 mm), 6 to 9 mu m (#30 file, 27 mm), and 5 to 7 mu m (#30, 31 mm) for all power settings. Antinodes occurring along the remaining file length were significantly larger at generator power 1 than at powers 2 through 5 (p < 0.03). Conclusions: At higher generator powers, energy delivered to the file is dissipated in unwanted vibration resulting in reduced vibration displacement amplitudes. This may reduce the occurrence of the biophysical forces necessary to maximize the technique's effectiveness. (J Endod 2010;36:880-883)

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据