4.5 Article

Evaluating Two Techniques on Removal of Fractured Rotary Nickel-Titanium Endodontic Instruments from Root Canals: An In Vitro Study

期刊

JOURNAL OF ENDODONTICS
卷 35, 期 4, 页码 559-562

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/j.joen.2008.12.019

关键词

Instrument removal; operating microscope; separated instrument; ultrasonics

向作者/读者索取更多资源

This study evaluated the ultrasonic technique (US) and instrument removal system (iRS) in removing fractured rotary nickel-titanium (NiTi) from root canals. A total of 30 extracted human molars with closed apices were collected. Profile .06 taper rotary endodontic instruments (Dentsply Maillefer, Ballaigues, Switzerland) 25-mm long and ISO size 25 were driven into the mesial canal to be fractured in the curve. Angles and radii of curvature were measured for all the canals with the fractured fragments, and then canals were classified into slight, moderate, and severe subgroups according to their curvature angles. Within each subgroup, teeth were randomly distributed into LIS or iRS according to measured angles and radii; a total of 15 canals were assigned to both US and iRS techniques. The overall success rate was 70% (n = 21). Less Curved canals and longer radii of curvature showed more success. The median time used for retrieval was 40 minutes using the US technique, whereas it was 55 minutes for iRS. The removal of fractured rotary NiTi endodontic instruments was more successful with less curved canals and longer radius of curvature above 4.4 mm. (J Endod 2009;35:559-562)