JOURNAL OF ELECTROCARDIOLOGY
卷 47, 期 6, 页码 881-889出版社
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1016/j.jelectrocard.2014.08.012
关键词
Low power; ECG; Holter; ECG patch; Full disclosure
Background: Detection of intermittent atrial fibrillation (AF) is done using a 24-h Holter. Halter recordings are powerful but lack the comfort and have limited recording times resulting in under diagnosing of intermittent AF. Objective: Within this work we evaluated and compared a novel miniaturized three-channel ECG monitoring patch versus a 24-h Halter system. Methods: Both patients with a chronic AF rhythm (n = 5) as well as patients with an AF rhythm that underwent electrical reconversion (n = 5) were equipped with both a 24-h Holter and ECG patch. Results: Alignment of raw data of both ECG systems allowed cross-correlation analysis. Overall good correlations of up to 85% were obtained. RR-interval analysis of both systems resulted in very high correlations of 99% and higher. AF analysis showed correct identification of AF on both ECG systems. Conclusions: The performance of our ECG patch matches that of the 24-h Holier and could provide a suitable tool for long-term monitoring applications. (C) 2014 Elsevier Inc. All rights reserved.