4.4 Article Proceedings Paper

First dose-map measured with a polycrystalline diamond 2D dosimeter under an intensity modulated radiotherapy beam

出版社

ELSEVIER
DOI: 10.1016/j.nima.2015.03.023

关键词

Diamond dosimctry; IMRT; Bidimensional dosimeters

资金

  1. INFN CSN5 DIAPIX
  2. IRPT/MIUR

向作者/读者索取更多资源

A prototype of bidimensional dosimeter made on a 2.5 x 2.5 cm(2) active area polycrystalline Chemical Vapour Deposited (pCVD) diamond film, equipped with a matrix of 12 x 12 contacts connected to the read-out electronics, has been used to evaluate a map of dose under Intensity Modulated Radiation Therapy (IMRT) fields for a possible application in pre-treatment verifications of cancer treatments. Tests have been performed under a 6-10 MVRX beams with IMRT fields for prostate and breast cancer. Measurements have been Laken by measuring the 144 pixels in different positions, obtained by shifting the device along the xly axes to span a total map of 14.4 x 10 cm(2). Results show that absorbed doses measured by our pCVD diamond device are consistent with those calculated by the Treatment Planning System (TPS). (C) 2015 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据