4.5 Article

Quantitative determination of atenolol in dried blood spot samples by LC-HRMS: A potential method for assessing medication adherence

出版社

ELSEVIER
DOI: 10.1016/j.jchromb.2012.04.013

关键词

Dried blood spot (DBS); Atenolol; Accurate mass; Guthrie card; LC-HRMS; Adherence

向作者/读者索取更多资源

The use of blood spot collection cards was investigated as a means of obtaining small volume samples for the quantification of therapeutic drugs for assessing medication adherence. A liquid chromatography-high resolution TOF mass spectrometry (LC-HRMS) method, based on the measurement at the accurate mass to charge ratio of the target analyte, was used to ensure specificity for atenolol in the dried blood spot (DBS) samples. A working method was developed and validated. For the preparation of DBS samples whole blood spiked with analyte was used to produce 30 mu l blood spots on specimen collection cards. A 5 mm disc was cut from the dried blood spot and extracted using methanol:water (60:40, v/v) containing the internal standard, atenolol-d(7). Extracts were vortexed, sonicated and then centrifuged. Gradient chromatographic elution was achieved using an Ascentis Express C18 100 mm x 2.1 mm column and a mobile phase flow rate of 0.2 ml/min and the column oven temperature at 30 degrees C. MS detection was carried out in electrospray positive ion mode for target ions at accurate mass m/z 267.1703 for atenolol and 274.2143 for the IS. Drug extraction efficiency from spiked blood spots was demonstrated to be 96 5% and the drug was stable in DBS for at least 10 weeks. The developed LC-HRMS method was linear within the tested calibration range of 25-1500 ng/ml and validation showed the accuracy (relative error) and precision (coefficient of variation) values were within the pre-defined limits of <= 15% at all concentrations with a limit of quantification of 25 ng/ml. Factors with potential to affect drug quantification measurements such as the matrix effects, volume of blood applied onto the collection card and effect of different sampling cards were investigated. The developed LC-HRMS method was applied to blood spots on sampling card taken from adult healthy volunteers previously administered a 50 mg atenolol tablet and a DBS concentration-time profile was obtained for atenolol. Requiring only a micro volume (30 mu l) blood sample for analysis, the developed DBS based assay has the potential to assess patient adherence to atenolol. (C) 2012 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据