4.5 Article

Full-scan high resolution accurate mass spectrometry (HRMS) in regulated bioanalysis: LC-HRMS for the quantitation of prednisone and prednisolone in human plasma

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.jchromb.2011.08.025

关键词

High resolution accurate mass spectrometry; Prednisone and prednisolone; Regulated bioanalysis; LC-MS/MS; LC-HRMS

向作者/读者索取更多资源

A liquid chromatography-full scan high resolution accurate mass spectrometry (LC-HRMS) method for quantifying prednisone and prednisolone in human plasma using a quadrupole time-of-flight mass spectrometer (Q-TOF) was developed. Plasma samples were extracted using a liquid-liquid extraction procedure. Full scan data were acquired in the TOF only mode and extracted ion chromatograms were generated post-acquisition with the exact masses of the analytes. The calibration range was 5-2500 ng/mL, with a Lower Limit of Quantitation (LLOQ) of 5 ng/mL. The assay accuracy was between 98.4% and 106.3%. The between-run (inter-day) and within-run (intra-day) precision were within 1.7% and 2.9%, respectively. The matrix effect was between 0.98 and 1.10 for the six different lots of human plasma evaluated. Pooled incurred samples were analyzed by the method and the results matched those obtained from an LC-MS/MS method. In addition, qualitative information on phospholipids, and other endogenous components were also extracted from the full-scan data acquired. (C) 2011 Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据